How to Set Up a Piston Cylinder Experiment
Note that web-based summaries of standard operating
procedures are not exhaustive.
Not all of the relevant safety information is included in this document.
Further training is required for safe operation and will be provided
in person.
|
Reserve a piston cylinder for the duration of your experiment using the online
calendar
The Basics: Terminology
|
|
|
|
|
|
|
| Stack |
Carbide Spacer |
Shim |
Piston |
Bridge |
Pressure Plate |
| |
|
|
|
|
|
|
Thermocouple Plate Insulator |
Thermocouple Plate |
Slotted Plate |
Large Spacer |
Upper Insulator |
Small Spacer |
|
|
|
|
|
|
|
Graphite Disk |
Cell Assembly |
Base Plug |
Thermocouple |
Sample (Lower) Ram |
Endload (Upper) Ram |
The Basics: Hydraulic System
The electric and hand operated hydraulic pumps are both connected to the same
hydraulic line. The valves A main and B main isolate each of the piston cylinders.
Turning the valve on the right side of
the hand pump clockwise prevents hydraulic fluid from draining back into the
tank and thus enables pressure generation with this pump. Likewise, when the
lever of the electric pump is pushed to the left, hydraulic fluid will not drain
back into the tank of the electric pump. Keep the hand pump valve closed
at all times other than when refilling the tank. In general, the electric pump
is used to close the daylight at the beginning of an experiment, and the hand
pump is used for controlled pressurisation. When the hydraulic cylinders are
retracting, the valve of the electric pump must be open
so that hydraulic fluid will return to the tank of this pump, which is the main
hydraulic reservoir for the system. You should always make sure that the tank
of the hand pump is filled before the start of an experiment. Close the main
valves that isolate PC A and PC B. Open the dust
cap to check the fluid level, close
the electric pump valve and open the hand
pump valve. Run the electric pump until the hand pump tank is full, close
the hand pump valve and replace the dust cap and turn to the vent
position.
Standard Operating Procedure
Preparing the standard cell assembly
- Make sure that all of the MgO pieces fit into the graphite furnace and that
the alumina thermocouple sleeve fits through both the base plug and the MgO.
- Drill a hole in the 19mm piece of MgO to fit your capsule. Use a carbide
end mill or a diamond drill to make a flat bottomed hole.
- Fire the MgO pieces at 1000 °C for ~12 h.
- Make the thermocouple. Bend the wires slowly using flat nosed pliers. The
wires are very brittle, so be careful not to crack them. Use Duco cement to
seal the holes at the top end of the insulator tube.

- Take the MgO out of the furnace. Allow it to cool under vacuum in a desiccator
for 15 min. The MgO should not be taken out of the desiccator until the assembly
is ready to load.

Building the stack
- Locate the carbide spacer
in the centre of the lower ram and place a shim
on top of the spacer.
- Place the bridge on top of the
spacer without displacing the shim. Make sure that the surface of the bridge
is clean and that the o-rings are in place. If necessary, remove the o-ring,
wipe it clean and apply vacuum grease before returning it to its groove.
- Clean the piston and insert
it into the hole in the bridge so that it rests on the shim.
- Clean the pressure plate.
Look carefully for cracks in the core and make a note of the number and nature
of the cracks.
- Lubricate the core of the pressure plate with moly-lube (thinnest possible
coating) and place the pressure plate over the piston.
- Push the graphite disk
in to the core.
- Take the cell components out of the vacuum desiccator. Insert the space
filling MgO, the capsule, the 1mm disk and the 12mm MgO with the hole for
the thermocouple tube into the furnace. Slide the furnace into the BaCO3 sleeve.
- Roll some lead foil to flatten and smooth it, and cut a piece of the correct
size to wrap around the BaCO3 without overlapping.

- Place the furnace inside the BaCO3 sleeve. Wrap the BaCO3 with Pb-foil.
The assembly goes inside the pressure vessel. Put the base
plug and the Pyrex sleeve on top of the cell and the push it all the way
down.
- If the top of the base plug is not flush with the top of the pressure plate,
close the valve of the electric pump and the endload (sample) ram, open the
main valve of your piston cylinder and close that of the other piston cylinder
and open the sample (lower) ram valve. Pump very slowly with the hand pump
until the base plug rises to the top of the pressure plate. Close the sample
ram valve.
- Put the thermocouple plate insulator
on top of the base plug. The hole of the mylar should be perfectly
aligned with the base plug. A small amount of silicone grease can be used
to make sure that the insulator stays in place.
- Insert the thermocouple
through the hole in the thermocouple plate, make sure that it goes all the
way in and is touching the 1mm MgO disk. Record the length of thermocouple
sticking out above the thermocouple plate (should be ~10 mm). The wires should
exit the thermocouple plate pointing to the side or the rear (so
that
in case of catastrophic failure you won't be hit in the eye by ejecta).
- Place the slotted plate
over the thermocouple plate. Align the slot toward the side.
- The large spacer with the
power cable attachment goes on top of the slotted plate.
- The plastic insulator and
the smaller steel plate (PC
B only) sit on top of the large spacer.
- You should now be ready to close the daylight and pressurize the stack.
Before you continue, pause to check that everything is in position. There
should be two pieces of insulating plastic in the correct
positions within the stack. Connect
the thermocouple wires and make sure that the thermocouple junction is closed.

Increasing pressure...
- Open the MAIN valve for your piston cylinder, and close that of the other.
Close the sample (lower) ram valve, open the endload (upper) ram valve. Run
the electric pump to close the daylight before using the hand pump to increase
the pressure to about 1000 psi. Close the endload (upper) ram valve.
- Attach the water hoses. The short hose connects the bridge to the thermocouple
plate. The two main hoses in the back should be attached to the thermocouple
plate and the bridge. Follow the color code: water enters at the bottom and
leaves at the top. Give the hoses a tug to make sure they're properly attached.
- Attach the power cable. Make sure the connection is tight.
- Make sure that the drain valves are closed. Turn the water on. Open the
water out valve first, followed by the water in valve.
- Feel the water flowing through the hoses -- oooh.
- If the low level warning light comes on, refill the chiller with coolant.
- Pressurize slowly and carefully. Each time you close the sample ram valve,
pump a little to increase the pressure in the hydraulic line before slowly
opening the endload valve, so that fluid is always bleeds in to the cylinder
and pressure increases as the valve is opened. The target pressure should
include a correction for friction.
Starting the experiment
Quenching the experiment
- Press the up and down arrow keys on the front panel of the Eurotherm controller
together. Immediately turn the potentiometer back to
0 and the heat off.
- When the thermocouple reads room temperature click the stop button on the
data logger.
- Close the water in and water out valves to stop the flow of water to the
apparatus.
- To drain water from the system, first open the drain valve, then slowly
open the air valve so that the water is flushed through the system and into
the drain. Once all the water is evacuated, close the air valve and then
close the drain valve. READ THIS!
- You've probably blown some of the water out of the trap, so add a little
water to prevent noxious gases from seeping back through the sewer pipes.
- Remove the water hoses from the stack.
- Disconnect the power cable from the stack.
- Disconnect the thermocouple.
Reducing the pressure
- The upper and lower ram valves should be closed at this point.
- Open the valve of the
electric pump.
- Very carefully apply some torque to open the sample (lower) ram valve, so
that the pressure slowly decreases. Do this as gently as possible - be patient.
- As the sample pressure decreases you may start to drain off some of the
endload, but be sure to do this carefully. Sample load must always be decreased
before endload.
- When the pressure is reduced and the hydraulic cylinders start to retract
you can disassemble the stack.
- After removing the slotted plate, break the thermocouple insulator and cut
the thermocouple wires. Return the thermocouple spool to its home.
- Remove the thermocouple plate. Wipe it dry and clean it.
- Remove the pressure plate. Cut the thermocouple wire that is sticking out
of the base plug, and clean the surface of any thermocouple debris.
- Dry and clean the bridge.
- The cell assembly and the piston should be pushed out of the pressure plate
using the 10 ton press. Before pushing out make sure that the push pin is
properly aligned with the base plug. The push pin MUST NOT OVERLAP any part
of the core.
- Close the shield and push out the assembly.
- Clean the pressure plate, check for cracks, and make notes of the observation.
Spray the pressure vessel with T9 and wipe it with a cloth to protect against
corrosion.
- Clean the piston and return it to its home.
- Start another experiment.
To prime the pressure controller
When you train someone to use the piston cylinder, read this document and
update any part of the procedure as required.
ACW 2005
The views and opinions expressed in this page
are strictly those of the page author.
The contents of this page have been neither reviewed nor approved by the University
of Minnesota.