-
[Skip this step if the power is already on] Turn on
the power to pressure and temperature controls by pulling on the power
switches.
-
Turn HEAT off.
-
Start Igor Pro and initialize the data logger (Logger
> Initialise Data Logger). If the logger fails to initialise, close Igor
and try disconnecting and reconnecting the USB-serial converter before starting
Igor again.
-
If the pressure controller is not fully retracted (decrease
pressure limit isn’t illuminated), reverse the motor to prime the
controller.
-
Remove wedges and examine thin plastic on wedges and
thicker plastic lining of the containment ring for damage. Replace if necessary.
When replacing lining there must be no overlap (put the new piece in, mark
it with a sharpie, take it out and cut off any excess with a scalpel). Wipe
the inside of the hatbox (inner surface of steel containment ring) with
a cloth moistened with T-9, to stop it rusting. Thin plastic must be glued
to the wedges with thinnest possible layer of loctite 495.
-
Vacuum dust from hatbox. DO NOT use compressed air to
blow dust around – it’ll get everywhere.
-
Clean surfaces using solvent (ethanol, methanol, whatever).
Lubricate inner surface of plastic lining containment ring and outer surfaces
of wedges with PTFE aerosol. Insert bottom wedges and G10 spacers.
-
Lubricate the assembled cube with PTFE.
-
Nest the assembled cube between the wedges. Insert type
K wires beneath G10 to make contact with an anvil. Attach thermocouple wires:
Type D
(+) : W97Re3, white insulation – compensated alloy 203
(-) : W75Re25, red insulation – compensated alloy 225
Type K
(+) : yellow insulation – Chromel
(-) : red insulation – Alumel
Ensure that there is no possibility of any wires shorting top and bottom
wedges.
-
Place the top wedges over the cube and drop the G10
spacers into the spaces between the wedges. Rest the top plate over the
center of
the wedges, and carefully roll the Walker module over the hydraulic cylinder.
Always make sure that the module rolls very slowly on the rails.
-
Mate thermocouple connectors and check for sensible temperature
reading.
-
Set up the data logger. Logger > Reset Data Logger.
Type in an experiment name and number. Click green start button and make
sure pressure is being monitored.
-
Set up the pressure-temperature program.
-
Open the compressed air valve. Open the path between
the pressure controller,
pump
and
multi anvil
(turn
valves
gently!). Press the 'close'
button to start the hydraulic pump and reduce the daylight.
As the ram closes, close
down the pump valve so that the pressure does not rise too quickly. As
the top plate makes contact with the overlying platen, the pump valve
should
be nearly closed. Watch
the pressure reading on the computer screen and adjust the valve to increase
pressure at about 1 psi per second. Continue to pressurize to ~200 psi.
At
>200 psi, slowly close the pump valve and turn off the pump. Close the
compressed air valve.

-
In data logger window select Logger > Motor
Control and manually increase the pressure controller motor speed (this
can be done before turning off the pump in the previous step if you want
to make a smooth transition between pressurizing systems). If the pressure-failure
alarm trips due to fluctuations in the pressure reading,
you will have to increase speed from zero again. Pressure controller and
multi anvil valves should be open a little. If line pressure drops much
below 200 psi, the pressure-failure alarm will trip continuously.
-
Start recording data (you can shift-click to start all
channels simultaneously). Start the pressure-temperature program cycle.
-
Check that cooling water hose is connected, the water
valves open. and the chiller on.
-
Check that OP is zero, hit the START/RESET
button and turn HEAT on.
-
Make sure the experiment record sheet is filed in the
ring binder, and that you’ve filled in your name, a description of
the experiment and the target conditions.
ACW 2007; updated 2012.